The Dwyer
Series 608 Differential Pressure Transmitters
convert positive, negative (vacuum), or differential pressures
of clean, dry air or other non-conductive, non-corrosive
gases into a standard two wire, 4-20 mA output signal. The
use of an ultra thin silicon diaphragm enables precision
measurement of differential pressures as low as 0.1" of
w.c. while withstanding high static working pressures up
to 100 psig (6.89 bar). The Series 608 transmitters are
FM approved intrinsically safe for use in the specified
hazardous locations when used with an approved intrinsic
safety barrier. The rugged NEMA 4X, stainless steel housing
makes this transmitter ideal for use in industrial and process
plant environments.
APPLICATIONS
- Lab Fume Hood Control
- Clean Room Applications
- Flow Measurements & Controls
- Filter Monitoring
- Furnace Draft Measurement
- Process Control
SPECIFICATIONS
Service: Clean/dry air and compatible gases.
Wetted Materials: Consult factory.
Accuracy: ±0.5% or ±0.25% full scale.
Stability: ±0.5% F.S./yr.
Pressure Limits: 100 psig (6.89 bar); 15 psid (1.03
bar).
Temperature Limits: -20 to 185°F (-28 to 85°C).
Compensated Temperature Range: 0 to 160°F (-18 to
71°C).
Thermal Effect: 0.5% Accuracy: ±0.02% F.S./°F; 0.25%
Accuracy: ±0.01% F.S./°F.
Power Requirements: 12 to 36 VDC (2-wire).
Output Signal: 4 to 20 mA DC.
Zero and Span Adjustments: Potentiometers for zero
and span.
Response Time: 250 ms.
Loop Resistance: DC: 0-1045 ohms maximum.
Current Consumption: 4-20 mA.
Electrical Connections: Screw terminals: two 1/2"
female NPT conduit.
Process Connections: Two 1/4" female NPT.
Enclosure Rating: NEMA 4X (IP65).
Mounting Orientation: Not position sensitive.
Weight: 2 lb (0.9 kg).
Agency Approvals: FM approved intrinsically safe
for use in Class I, Div. 1, Groups A, B, C, D; Class II,
Div. 1, Groups E, F, G; Class III, Div. 1 when wire with
approved intrinsically safe barrier. Entity parameters:
Vmax=36 VDC; Imax=250 mA; C=12 nF; L=0 mH. |