Low cost,
durable sanitary pressure transmitter is designed
to meet 3A standards allowing use in food, dairy, beverage
and pharmaceutical industries. This unit features superior
±0.25% accuracy and 5:1 turndown. Calibration is quick and
easy with field accessible zero and span adjustments. All
stainless steel construction resists the corrosive effects
of caustic washes used in most food processing cleaning
procedures. 1-1/2" Sanitary Clamp process connection is
suitable for Clean-in-Place (CIP) applications, eliminating
the need for expensive bypass piping, valving or removal
requirements for steam cleaning.
SPECIFICATIONS
Service: Liquid, gas or vapor.
Wetted Materials: 316L stainless steel.
Body: 316 stainless steel.
Accuracy: ±0.25% of calibrated span.
Temperature Limits: Process Interface -40 to 212°F
(-40 to 100°C).
Pressure Limits: 300% full scale.
Compensated Temperature Range: -20 to 180°F (-25
to 80°C).
Thermal Effect: (includes zero and span) ±1% of upper
range limit per 50°F (30 to 130°F); ±1.6% of upper range
limit per 50°F (10 to 180°F).
Power Requirements: 12-40 VDC with reverse polarity
protection.
Output Signal: 4-20 mA DC, max. 30 mA DC (2 wire).
Zero and Span Adjustments: ±10% each.
Response Time: Time constant, 20 ms.
Loop Resistance: 600 ohms @ 24 VDC; max. ohms = (supply
voltage - 12) x 50.
Electrical Connection: 1/2" female NPT.
Process Connection: 1-1/2" Sanitary Clamp.
Weight: 1.67 lb (752 g).
Agency Approvals: CE.
|