Model 660
Air Velocity Monitors are a practical, affordable
way to continuously monitor for safe airflows through laboratory
fume hoods. They are typically installed in the fume hood
side fascia and connected to the interior sidewall via 1-1/8"
flexible tubing. As the exhaust fan draws air through the
device, a sensitive constant temperature thermistor measures
flow and lights a green (normal), yellow (high) or red (low)
LED. An audible alarm also warns of low flow and requires
manual resetting. Mounting holes fit standard single gang
electrical box.SPECIFICATIONS
Service: Air and non-combustible,
non-corrosive gases.
High Setpoint Range: 0-275 FPM (0-1.397 m/s).
Low Setpoint Range: 0-150 FPM (0-.792 m/s).
Repeatability: ±7% of full span, 0-50 and 150-275 FPM;
±5% of full span, 50-150 FPM.
Compensated Temperature Range: 50 to 90°F (10 to 32.2°C).
Operating Temperature Range: 32 to 120°F (0 to 48.9°C).
Power Supply: 24 VAC, 3 watts maximum.
Response Time: 6-10 seconds.
Warm-Up Time: 3-5 minutes (no flow).
Audible Alarm: 75 dB between 3-10 feet.
Weight: 3 oz (85 g).
Mounting Hardware: (2) #6 x 1" sheet metal screws,
(2) 6-32 x 7/8" machine screws.
|