Calibrate
electronic process instruments such as controllers, transmitters,
or recorders with the model CAL30 Process Calibrator. Model
CAL30 is designed to independently measure and simulate
DC volts, mV, and current. These calibrators simulate a
two-wire field transmitter and generate a controlled output
current when connected to an external power supply. Units
source is a four-wire transmitter by generating a controlled
4-20 mA DC from an external 25 V power supply.
Programmable functions, such as square root
extraction for flow signals, ramp/step increment value,
selectable engineering units, isolated signal conversion,
or automatic changing of the output value simplify many
calibration tasks. Calibration values can be stored and
recalled using the Low, Mid, or High keys. Program the automatic
shut-off to conserve battery life.
The Model CAL30 Calibrator features 500 VDC
input/output isolation and reverse polarity protection.
The unit includes three 9V batteries, two sets of dual test
leads, a dual binding post adapter, carrying case, and instruction
manual.
SPECIFICATIONS
Input Ranges: Voltage: -10 to 24 mV DC; -10 to 120
mV DC (autoranging); 0 to 12 VDC; Current: 0 to 24 mA DC.
Input Impedance: 1 MO minimum plus 10 nano-Amps maximum
for V and mV inputs; 40O for mA DC.
Accuracy: ±0.025% of range.
Output Ranges: 0 to 120.0 mV DC (2KO min. load);
0 to 12.0 VDC (2KO min. load); 0 to 24.0 mA DC (25 volts
excitation).
Input/Output Isolation: 500 VDC.
Display: 24-character LCD, 3/16" (5 mm) character
height.
Indication: Up to 5 digits depending on range.
Warm-up Time: 60 seconds, maximum.
Ambient Operating Temperature: -4 to 113°F (-20 to
45°C).
Power Requirements: Three 9V alkaline (included).
Battery Life: 100 hours, approx.
Weight: 1.15 lb (0.52 kg).
Agency Approvals: CE.
|