Model MC6
Multi-Cal Benchtop Calibrator performs a wide
variety of simple and complex pressure based measurement,
test, and calibration operations. Modular sensor design
allows user to select pressure measurement range for application
flexibility. Calibrator can accommodate up to two interchangeable
pressure modules (sold separately) in any combination of
range or accuracy.
Simultaneously display two separate measurements
on the two line, alphanumeric display. Readings can be displayed
in a choice of twelve preprogrammed engineering units or
any single user-defined unit. Calibrator features min/max
recall, hi/lo alarm, percentage of full scale pressure readings,
mA/voltage measurement, leak rate and pressure decay measurement,
switch testing capabilities, including trip point and dead
band, and velocity/volume flow rates.
Quickly document calibration procedures using
the data logging feature which stores up to 384 sets of
pressure and time/date stamped measurements. The automatic
% error calculation simplifies transmitter calibration.
Easily upload stored data to an IBM compatible computer
via the RS232 two-way communications port. Calibrators and
modules include certification to NIST traceability--ideal
as a secondary standard for calibrating pressure equipment.
Multi-Cal Pressure Calibrator includes utility
software, protection module, adjustment handle, AC adapter,
and instruction manual.
Multi-Cal
Pressure Modules are interchangeable and available
in a wide selection of pressure ranges and accuracies. Benchtop
calibrator accepts up to two pressure modules. Modules include
NIST calibration certification.
APPLICATIONS
Pressure measurement for clean room control, filter performance
monitoring, HVAC testing and setup, draft measurement, airflow
measurement control, differential pressure measurement in
laminar flowhoods, paint booths, industrial ovens and fume
hoods. Use as a secondary standard for calibrating pressure
equipment.
SPECIFICATIONS
Service: Clean, dry, nonconductive, noncorrosive
gases.
Accuracy: Differential Pressure Modules: ±0.06% FS,
Gauge Pressure Modules: ±0.05% FS, Voltage Input: ±0.025%
FS @ 1/10 VDC, ±0.10% FS @ 0/30 VDC, Current Input: ±0.03%
FS @ 1/20 mA, ±0.05% FS @ 0/50 mA.
Sensitivity: ±0.002% of span with dampening 1 part
in 50,000 (max.).
Repeatability: Ranges =0/2 psi: ±0.05% of span, Ranges
=0/5 psid: ±0.02% of span.
Output: RS232 serial interface, 9-pin.
Alarm Output: SPST form C 110 VDC, 120 VDC (max.),
1 A (mas.), 30 W, 62.5 VA (resistive).
Display: Alphanumeric LCD, 0.37" (9.5 mm) height
per line, 2 lines, 16 characters/line.
Display Update: 100 msec.
Ambient Operation Temperature: 32 to 120°F (0 to
49°C).
Storage Temperature: -4 to 158°F (-20 to 70°C).
Electrical Connections: Starndard banana jacks.
Power Supply: Internal: two 9 VDC alkaline batteries,External:
AC adapter 9 VDC, 500 mA.
Battery Life: 30 hours (approximate).
Engineering Units: in H2O, psi, in Hg, kPa, mbar, bar, cm H2O, mm Hg, ft SW, MPa, mm H2O, kg cm²
and user-defined.
Overpressure: Differential Pressure Modules: 50 psi
positive direction, 15 psi negative direction, Gauge Pressure
Modules: 2x range (0/5 psi to 0/1000 psi).
Temperature Compensation: 20 to 120°F (-7 to 49°C).
Temperature Error: Maximum of ±0.004% of span per
°F over compensated range for zero and span.
Temperature Effect Electrical Measurement: ±0.001%
of span per °F over compensated range.
Dampening: (Measurement averaging) programmable from
0 to 16 consecutive readings.
Baud Rate: 300, 1200, 2400, or 9600, selectable.
Computer Requirements: IBM compatible 386 or above
and Windows® 3.1 or later with 2mB RAM and 2mB hard drive
disk space, one serial port.
Housing Material: ABS plastic.
Weight: Calibrator: 3 lb (1.4 kg), Pressure Module:
0.5 lb (0.3 kg).
Agency Approvals: CE.
OPTIONS
-- Consult factory
For other pressure ranges and accuracies.