Accurately
measure flow rates of air, water, and other commonly
used gases with the Series VA Variable Area Glass Flowmeters.
Flowmeters are designed with an easy-to-read universal mm
scale and supplied with a correlation chart containing calibration
data for air and water. Correlation data for other gases
and liquids are available. Permanently fused ceramic scale
with vertical locator line reduces parallax and eye fatigue.
Thick polycarbonate front shield protects tube from breakage
and also serves as a magnifying lens to enhance reading
resolution.
Glass flowmeters are suitable for metering
carrier gases, liquid and gas measurement in pilot plants,
laboratories, process flow and level indicating. Units are
equipped with a standard 6-turn needle valve for flow rate
control. High precision metering valves with non-rising
stems (sold separately) are available for high sensitivity
control and resolution - particularly suited for very low
flow rates.
Flowmeters are shipped completely assembled
and ready for panel mounting. Use an optional acrylic tripod
base to convert to self-standing bench mounting.
SPECIFICATIONS
Services: Compatible gases or liquids.
Wetted Materials:
Flowtube: Borosilicate glass.
Floats: Glass or stainless steel (sapphire, Carboloy and
tantalum are optional).
Float Stops: PTFE.
End Fittings: Anodized Aluminum, 316 Stainless Steel or
Brass.
Packings: Fluoroelastomer.
O-rings: Buna-N on aluminum models and brass models, and
Fluoroelastomer on stainless steel models.
Temperature Limits: 250°F (121°C).
Pressure Limits: 200 psig (13.8 bar).
Accuracy: ±2% FS @ 70°F (21.1°C) and 14.7 psia (1
atm absolute).
Repeatability: ±0.25% full scale.
Scales: Universal 65 mm or 150 mm with correlation
charts.
Turn-down Ratio: 10:1; 20:1 with combinations of
two floats installed in meter.
Connections: Two 1/8" female NPT.
Mounting: Vertical.
Valve: 6-turn needle (standard), optional 16-turn
high precision valve.
Valve Orifice: Acetal on aluminum models and brass
models, PCTFE on stainless steel models.
|