Measure flow of corrosive or ultra-pure liquids
or gases with the Series VA Variable Area PTFE/Glass Flowmeters.
Flowmeters are constructed of chemically inert materials
and housed in a rigid anodized aluminum frame with a polycarbonate
safety shield. Units are designed with an easy-to-read universal
mm scale and supplied with a correlation chart containing
calibration data for air and water. Correlation data for
other gases and liquids are available.
SPECIFICATIONS
Service: Compatible gases or liquids.
Wetted Materials:
Flowtube: Borosilicate glass.
Floats: Glass sapphire optional).
Float Stops and End Fittings: PTFE.
O-rings and Packings: PTFE.
Temperature Limits: -15 to to 150°F (-26 to 65°C).
Pressure Limits: 100 psig (6.7 bar).
Accuracy: ±2% FS @ 70°F (21.1°C) and 14.7 psia (1
atm absolute).
Repeatability: ±0.25%.
Leak Integrity: 1 x 10-7 sccs of Helium.
Scales: Universal 65 mm or 150 mm with correlation
charts.
Turn-down ratio: 10:1, 20:1 with combinations of
two floats installed in meter.
Connections: Two 1/8" female NPT.
Mounting: Vertical.
Valve: 6-turn needle.
Valve Orifice: PTFE. |